Oxycodone Hydrochloride
(ox'' i koe' done hye'' droe klor' ide).
C18H21NO4·HCl 351.82 Morphinan-6-one, 4,5-epoxy-14-hydroxy-3-methoxy-17-methyl-, hydrochloride, (5)-; 4,5-Epoxy-14-hydroxy-3-methoxy-17-methylmorphinan-6-one hydrochloride [124-90-3]. DEFINITION
Oxycodone Hydrochloride contains NLT 97.0% and NMT 103.0% of C18H21NO4·HCl, calculated on the anhydrous, solvent-free basis.
IDENTIFICATION
• A. Procedure
Sample solution:
Dissolve 250 mg in 25 mL of water.
Analysis:
Render the 25 mL of Sample solution alkaline with 6 N ammonium hydroxide. Allow the mixture to stand until a precipitate is formed. Filter, wash the precipitate with 50 mL of cold water, and dry for 2 h at 105.
Acceptance criteria:
The precipitate so obtained melts between 218 and 223, but the range between the beginning and the end of melting does not exceed 2 (see Melting Range or Temperature 741).
• B. Infrared Absorption 197K:
Use a portion of the dried precipitate obtained in Identification test A.
ASSAY
• Procedure
Mobile phase:
0.005 M sodium 1-hexanesulfonate, methanol, triethylamine, and phosphoric acid (900:100:2:5). Adjust with 50% sodium hydroxide solution to a pH of 2.5 ± 0.1, and filter.
System suitability solution:
13 µg/mL of codeine phosphate and 9 µg/mL of oxycodone in Mobile phase
Standard solution:
0.9 mg/mL of USP Oxycodone RS in Mobile phase
Sample solution:
1 mg/mL of Oxycodone Hydrochloride in Mobile phase. [NotePass a portion of this solution through a filter having a 0.5-µm or finer pore size, and use the filtrate as the Sample solution. ]
Chromatographic system
Mode:
LC
Detector:
UV 206 nm
Column:
3.9-mm × 15-cm; 4-µm packing L7
Column temperature:
50
Flow rate:
1.5 mL/min
Injection size:
10 µL
System suitability
Samples:
System suitability solution and Standard solution
[NoteThe relative retention times for codeine and oxycodone are about 0.8 and 1.0, respectively. ]
Suitability requirements
Resolution:
NLT 3.0 between codeine and oxycodone, System suitability solution
Tailing factor:
0.751.25, Standard solution
Relative standard deviation:
NMT 2.0% from replicate injections, Standard solution
Analysis
Samples:
Standard solution and Sample solution
[NoteRecord the chromatograms for a period of time that is twice the retention time of the main oxycodone peak. ]
Calculate the percentage of C18H21NO4·HCl in the portion of Oxycodone Hydrochloride taken:
Result = (rU/rS) × (CS/CU) × (Mr1/Mr2) × 100
Acceptance criteria:
97.0%103.0% on the anhydrous, solvent-free basis
IMPURITIES
Inorganic Impurities
• Residue on Ignition 281:
NMT 0.05%. [NoteUse of sulfuric acid is omitted. ]
Organic Impurities
• Procedure 1: Limit of Alcohol
Internal standard stock solution:
Transfer 6.0 mL of isopropyl alcohol to a 500-mL volumetric flask, and dilute with water to volume. [NoteThe isopropyl alcohol must be free of alcohol impurities. ]
Internal standard solution:
Transfer 5.0 mL of the Internal standard stock solution to a 100-mL volumetric flask, and dilute with water to volume.
Standard stock solution:
16 mg/mL of alcohol (C2H5OH) in water
Standard solution:
Pipet 3.0 mL of the Standard stock solution and 5.0 mL of the Internal standard stock solution into a 100-mL volumetric flask, and dilute with water to volume.
Sample solution:
Transfer about 240 mg of Oxycodone Hydrochloride to a 15-mL centrifuge tube, add 5.0 mL of the Internal standard solution, and mix to dissolve.
Chromatographic system
Mode:
GC
Detector:
Flame ionization
Column:
4-mm × 1.8-m glass; packed with 80- to 100-mesh support S3
Carrier gas:
Helium
Temperature
Column:
150. [NoteCondition the column overnight at 235 with a slow flow of carrier gas. ]
Injector:
170
Detector:
170
Injection size:
5 µL
System suitability
Sample:
Standard solution
Suitability requirements
Resolution:
NLT 2 between isopropyl alcohol and alcohol
Tailing factor:
NMT 1.5
Relative standard deviation:
NMT 2.0%
Analysis
Samples:
Standard solution and Sample solution
Calculate the percentage of alcohol (C2H5OH) in the portion of Oxycodone Hydrochloride taken:
Result = (RU/RS) × (CS/CU) × 100
Acceptance criteria:
NMT 1.0%
• Procedure 2
Analysis:
Use the chromatogram of the Sample solution in the Assay to calculate the percentage of each impurity in the portion of Oxycodone Hydrochloride taken:
Result = (rU/rT) × 100
Acceptance criteria
Individual impurities:
The impurities meet the requirements listed in Impurity Table 1.
Total impurities:
NMT 2.0%
Impurity Table 1
• Procedure 3: Limit of Oxycodone Related Compound A (14-Hydroxycodeinone) and Oxycodone Related Compound C (Codeinone)
Solution A:
Dissolve 3.45 g of monobasic sodium phosphate in 1000 mL of water. Add 5.41 g of sodium dodecyl sulfate, and mix. Filter, and adjust with 50% (w/v) sodium hydroxide solution to a pH of 7.50 ± 0.05.
Solution B:
Water and phosphoric acid (9:1)
Diluent:
Prepare a mixture of water and Solution B (9:1).
Mobile phase:
Prepare a mixture of acetonitrile, methanol, and Solution A (15.8:12.0:72.2), and adjust with Solution B to a pH of 7.80 ± 0.01.
Standard solution:
50 mg/mL of USP Oxycodone Hydrochloride RS and 0.5 µg/mL each of USP Oxycodone Related Compound A RS and USP Oxycodone Related Compound C RS in Diluent
Unspiked oxycodone hydrochloride solution:
50 mg/mL of USP Oxycodone Hydrochloride RS in Diluent
System suitability solution:
100 µg/mL of USP Oxycodone Hydrochloride RS and 5 µg/mL each of USP Oxycodone Related Compound A RS and USP Oxycodone Related Compound C RS in Diluent
Sample solution:
50 mg/mL of Oxycodone Hydrochloride in Diluent
Chromatographic system
Mode:
LC
Detector:
UV 220 nm
Column:
3.0-mm × 15-cm; 3.5-µm packing L1
Column temperature:
40
Flow rate:
0.7 mL/min
Injection size:
5 µL
System suitability
Samples:
Standard solution and System suitability solution
[NoteThe relative retention times for oxycodone related compound C, oxycodone related compound A, and oxycodone are about 0.44, about 0.85, and 1.0, respectively. ]
Suitability requirements
Resolution:
NLT 4 between oxycodone related compound A and oxycodone related compound C, System suitability solution
Tailing factor:
NMT 2.0, System suitability solution
Relative standard deviation:
NMT 20% for oxycodone related compound A and C, Standard solution
Analysis
Samples:
Diluent, Standard solution, Unspiked oxycodone hydrochloride solution, and Sample solution
Calculate the percentage of oxycodone related compound A and oxycodone related compound C in the portion of Oxycodone Hydrochloride taken:
Result = (rU/rS) × (CS/CU) × 100
Acceptance criteria
Individual impurities:
See Impurity Table 2.
(Procedure 3 postponed indefinitely)
SPECIFIC TESTS
• Content of Chloride
Sample solution:
6 mg/mL in methanol
Analysis:
To 50 mL of Sample solution add 5 mL of glacial acetic acid, and titrate with 0.1 N silver nitrate VS, determining the endpoint potentiometrically. Each mL of 0.1 N silver nitrate is equivalent to 3.545 mg of Cl.
Acceptance criteria:
9.8%10.4% calculated on the anhydrous, solvent-free basis
• Optical Rotation, Specific Rotation 781S:
137 to 149
Sample solution:
25 mg/mL of Oxycodone Hydrochloride in water, on the anhydrous, solvent-free basis
• Water Determination, Method I 921:
NMT 7.0%
ADDITIONAL REQUIREMENTS
• Packaging and Storage:
Preserve in tight containers.
• USP Reference Standards 11
USP Oxycodone Related Compound A RS
14-Hydroxycodeinone.
USP Oxycodone Related Compound C RS
Codeinone.
Auxiliary Information
Please check for your question in the FAQs before contacting USP.
USP35NF30 Page 4170
Pharmacopeial Forum: Volume No. 34(6) Page 1480
|