Oxycodone Hydrochloride
(ox'' i koe' done hye'' droe klor' ide).
Click to View Image

C18H21NO4·HCl 351.82
Morphinan-6-one, 4,5-epoxy-14-hydroxy-3-methoxy-17-methyl-, hydrochloride, (5)-;    
4,5-Epoxy-14-hydroxy-3-methoxy-17-methylmorphinan-6-one hydrochloride     [124-90-3].
DEFINITION
Oxycodone Hydrochloride contains NLT 97.0% and NMT 103.0% of C18H21NO4·HCl, calculated on the anhydrous, solvent-free basis.
IDENTIFICATION
•  A. Procedure
Sample solution:  Dissolve 250 mg in 25 mL of water.
Analysis:  Render the 25 mL of Sample solution alkaline with 6 N ammonium hydroxide. Allow the mixture to stand until a precipitate is formed. Filter, wash the precipitate with 50 mL of cold water, and dry for 2 h at 105.
Acceptance criteria:  The precipitate so obtained melts between 218 and 223, but the range between the beginning and the end of melting does not exceed 2 (see Melting Range or Temperature 741).
•  B. Infrared Absorption 197K: Use a portion of the dried precipitate obtained in Identification test A.
ASSAY
•  Procedure
Mobile phase:  0.005 M sodium 1-hexanesulfonate, methanol, triethylamine, and phosphoric acid (900:100:2:5). Adjust with 50% sodium hydroxide solution to a pH of 2.5 ± 0.1, and filter.
System suitability solution:  13 µg/mL of codeine phosphate and 9 µg/mL of oxycodone in Mobile phase
Standard solution:  0.9 mg/mL of USP Oxycodone RS in Mobile phase
Sample solution:  1 mg/mL of Oxycodone Hydrochloride in Mobile phase. [Note—Pass a portion of this solution through a filter having a 0.5-µm or finer pore size, and use the filtrate as the Sample solution. ]
Chromatographic system 
Mode:  LC
Detector:  UV 206 nm
Column:  3.9-mm × 15-cm; 4-µm packing L7
Column temperature:  50
Flow rate:  1.5 mL/min
Injection size:  10 µL
System suitability 
Samples:  System suitability solution and Standard solution
[Note—The relative retention times for codeine and oxycodone are about 0.8 and 1.0, respectively. ]
Suitability requirements 
Resolution:  NLT 3.0 between codeine and oxycodone, System suitability solution
Tailing factor:  0.75–1.25, Standard solution
Relative standard deviation:  NMT 2.0% from replicate injections, Standard solution
Analysis 
Samples:  Standard solution and Sample solution
[Note—Record the chromatograms for a period of time that is twice the retention time of the main oxycodone peak. ]
Calculate the percentage of C18H21NO4·HCl in the portion of Oxycodone Hydrochloride taken:
Result = (rU/rS) × (CS/CU) × (Mr1/Mr2) × 100
rU== peak response from the Sample solution
rS== peak response from the Standard solution
CS== concentration of USP Oxycodone RS in the Standard solution (mg/mL)
CU== concentration of Oxycodone Hydrochloride in the Sample solution (mg/mL)
Mr1== molecular weight of oxycodone hydrochloride, 351.82
Mr2== molecular weight of oxycodone base, 315.37
Acceptance criteria:  97.0%–103.0% on the anhydrous, solvent-free basis
IMPURITIES
Inorganic Impurities 
•  Residue on Ignition 281: NMT 0.05%. [Note—Use of sulfuric acid is omitted. ]
Organic Impurities 
•  Procedure 1: Limit of Alcohol
Internal standard stock solution:  Transfer 6.0 mL of isopropyl alcohol to a 500-mL volumetric flask, and dilute with water to volume. [Note—The isopropyl alcohol must be free of alcohol impurities. ]
Internal standard solution:  Transfer 5.0 mL of the Internal standard stock solution to a 100-mL volumetric flask, and dilute with water to volume.
Standard stock solution:  16 mg/mL of alcohol (C2H5OH) in water
Standard solution:  Pipet 3.0 mL of the Standard stock solution and 5.0 mL of the Internal standard stock solution into a 100-mL volumetric flask, and dilute with water to volume.
Sample solution:  Transfer about 240 mg of Oxycodone Hydrochloride to a 15-mL centrifuge tube, add 5.0 mL of the Internal standard solution, and mix to dissolve.
Chromatographic system 
Mode:  GC
Detector:  Flame ionization
Column:  4-mm × 1.8-m glass; packed with 80- to 100-mesh support S3
Carrier gas:  Helium
Temperature 
Column:  150. [Note—Condition the column overnight at 235 with a slow flow of carrier gas. ]
Injector:  170
Detector:  170
Injection size:  5 µL
System suitability 
Sample:  Standard solution
Suitability requirements 
Resolution:  NLT 2 between isopropyl alcohol and alcohol
Tailing factor:  NMT 1.5
Relative standard deviation:  NMT 2.0%
Analysis 
Samples:  Standard solution and Sample solution
Calculate the percentage of alcohol (C2H5OH) in the portion of Oxycodone Hydrochloride taken:
Result = (RU/RS) × (CS/CU) × 100
RU== peak response ratio of the alcohol peak to the isopropyl alcohol from the Sample solution
RS== peak response ratio of the alcohol peak to the isopropyl alcohol from the Standard solution
CS== concentration of alcohol in the Standard solution (mg/mL)
CU== concentration of Oxycodone Hydrochloride in the Sample solution (mg/mL)
Acceptance criteria:  NMT 1.0%
•  Procedure 2
Analysis:  Use the chromatogram of the Sample solution in the Assay to calculate the percentage of each impurity in the portion of Oxycodone Hydrochloride taken:
Result = (rU/rT) × 100
rU== peak response for each impurity
rT== sum of the responses of all the peaks
Acceptance criteria 
Individual impurities:  The impurities meet the requirements listed in Impurity Table 1.
Total impurities:  NMT 2.0%
Impurity Table 1
Name Relative Retention Time Acceptance
Criteria,
NMT (%)
Oxymorphone 0.31 0.15
Noroxymorphone 0.33 0.15
10-Hydroxyoxycodone 0.53 0.15
6- Oxycodol 0.67 0.25
7,8-Dihydro-8-14-dihydroxycodeinone 0.71 0.15
Hydrocodone 1.19 0.15
Individual unspecified impurity 0.10
•  Procedure 3: Limit of Oxycodone Related Compound A (14-Hydroxycodeinone) and Oxycodone Related Compound C (Codeinone)
Solution A:  Dissolve 3.45 g of monobasic sodium phosphate in 1000 mL of water. Add 5.41 g of sodium dodecyl sulfate, and mix. Filter, and adjust with 50% (w/v) sodium hydroxide solution to a pH of 7.50 ± 0.05.
Solution B:  Water and phosphoric acid (9:1)
Diluent:  Prepare a mixture of water and Solution B (9:1).
Mobile phase:  Prepare a mixture of acetonitrile, methanol, and Solution A (15.8:12.0:72.2), and adjust with Solution B to a pH of 7.80 ± 0.01.
Standard solution:  50 mg/mL of USP Oxycodone Hydrochloride RS and 0.5 µg/mL each of USP Oxycodone Related Compound A RS and USP Oxycodone Related Compound C RS in Diluent
Unspiked oxycodone hydrochloride solution:  50 mg/mL of USP Oxycodone Hydrochloride RS in Diluent
System suitability solution:  100 µg/mL of USP Oxycodone Hydrochloride RS and 5 µg/mL each of USP Oxycodone Related Compound A RS and USP Oxycodone Related Compound C RS in Diluent
Sample solution:  50 mg/mL of Oxycodone Hydrochloride in Diluent
Chromatographic system 
Mode:  LC
Detector:  UV 220 nm
Column:  3.0-mm × 15-cm; 3.5-µm packing L1
Column temperature:  40
Flow rate:  0.7 mL/min
Injection size:  5 µL
System suitability 
Samples:  Standard solution and System suitability solution
[Note—The relative retention times for oxycodone related compound C, oxycodone related compound A, and oxycodone are about 0.44, about 0.85, and 1.0, respectively. ]
Suitability requirements 
Resolution:  NLT 4 between oxycodone related compound A and oxycodone related compound C, System suitability solution
Tailing factor:  NMT 2.0, System suitability solution
Relative standard deviation:  NMT 20% for oxycodone related compound A and C, Standard solution
Analysis 
Samples:  Diluent, Standard solution, Unspiked oxycodone hydrochloride solution, and Sample solution
Calculate the percentage of oxycodone related compound A and oxycodone related compound C in the portion of Oxycodone Hydrochloride taken:
Result = (rU/rS) × (CS/CU) × 100
rU== peak response of oxycodone related compound A or oxycodone related compound C from the Sample solution
rS== peak response of oxycodone related compound A or oxycodone related compound C minus the response of the Unspiked oxycodone hydrochloride solution from the Standard solution
CS== concentration of USP Oxycodone Related Compound A RS or USP Oxycodone Related Compound C RS in the Standard solution (mg/mL)
CU== concentration of Oxycodone Hydrochloride in the Sample solution (mg/mL)
Acceptance criteria 
Individual impurities:  See Impurity Table 2.
Impurity Table 2
Name Relative
Retention
Time
Acceptance
Criteria,
NMT (%)
Oxycodone related compound Aa 0.85 0.001
Oxycodone related compound Cb 0.44 0.001
Oxycodone 1.0
a  14-Hydroxycodeinone (Cl8H21NO4).
b  Codeinone (C18H19NO3).
(Procedure 3 postponed indefinitely)
SPECIFIC TESTS
•  Content of Chloride
Sample solution:  6 mg/mL in methanol
Analysis:  To 50 mL of Sample solution add 5 mL of glacial acetic acid, and titrate with 0.1 N silver nitrate VS, determining the endpoint potentiometrically. Each mL of 0.1 N silver nitrate is equivalent to 3.545 mg of Cl.
Acceptance criteria:  9.8%–10.4% calculated on the anhydrous, solvent-free basis
•  Optical Rotation, Specific Rotation 781S: 137 to 149
Sample solution:  25 mg/mL of Oxycodone Hydrochloride in water, on the anhydrous, solvent-free basis
ADDITIONAL REQUIREMENTS
•  Packaging and Storage: Preserve in tight containers.
•  USP Reference Standards 11
USP Oxycodone RS Click to View Structure
USP Oxycodone Hydrochloride RS Click to View Structure
USP Oxycodone Related Compound A RS
14-Hydroxycodeinone.
USP Oxycodone Related Compound C RS
Codeinone.
Auxiliary Information— Please check for your question in the FAQs before contacting USP.
Topic/Question Contact Expert Committee
Monograph Clydewyn M. Anthony, Ph.D.
Senior Scientific Liaison
1-301-816-8139
(SM22010) Monographs - Small Molecules 2
Reference Standards RS Technical Services
1-301-816-8129
rstech@usp.org
USP35–NF30 Page 4170
Pharmacopeial Forum: Volume No. 34(6) Page 1480