Add the following:
Orlistat Capsules
DEFINITION
Orlistat Capsules contain NLT 90.0% and NMT 110.0% of the labeled amount of orlistat (C29H53NO5).
IDENTIFICATION
• The retention time of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.
ASSAY
• Procedure
Mobile phase:
Acetonitrile, phosphoric acid, and water (860: 0.05: 140)
Standard solution:
0.6 mg/mL of USP Orlistat RS in Mobile phase
Sample solution:
Transfer the contents of NLT 10 Capsules into a suitable container, weigh, and mix. Transfer an accurately weighed portion of the powder, equivalent to 120 mg of orlistat, into a 200-mL volumetric flask. Add 140 mL of Mobile phase, and sonicate for 1 min. Shake the resulting solution mechanically for 15 min, and dilute with Mobile phase to volume. Pass a portion of this solution through a filter of 0.45-µm or finer pore size, discarding the first few mL of filtrate.
Chromatographic system
Mode:
LC
Detector:
195 nm
Column:
3.9-mm × 15-cm; packing L1
Flow rate:
1.0 mL/min
Injection size:
20 µL
System suitability
Sample:
Standard solution
System suitability requirements
Relative standard deviation:
NMT 2.0%
Analysis
Samples:
Standard solution and Sample solution
Calculate the percentage of the labeled amount of orlistat (C29H53NO5) in the portion of Capsules taken:
Result = (rU/rS) × (CS/CU) × 100
Acceptance criteria:
90.0%110.0%
PERFORMANCE TESTS
• Dissolution 711
Medium:
3% Sodium lauryl sulfate and 0.5% sodium chloride in water. To each 10 L of media add 12 drops of n-octanol, and adjust with phosphoric acid to a pH of 6.0; 900 mL.
Apparatus 2:
75 rpm, with coil wire sinker
Time:
45 min
Mobile phase:
Acetonitrile and water (860:140)
Standard solution:
Transfer about 13 mg of USP Orlistat RS to a 100-mL volumetric flask. Dissolve in 2 mL of acetonitrile, and dilute with Medium to volume.
Sample solution:
Pass a portion of the solution under test through a suitable filter of 0.2-µm pore size.
Chromatographic system
Mode, Detector, and Column:
Proceed as directed in the Assay.
Flow rate:
2.0 mL/min
Injection size:
50 µL
System suitability
Sample:
Standard solution
System suitability requirements
Relative standard deviation:
NMT 2.0%
Calculate the percentage of the labeled amount of orlistat (C29H53NO5) dissolved:
Result = (rU/rS) × (CS/L) × V × 100
Tolerances:
NLT 75% (Q) of the labeled amount of orlistat is dissolved.
• Uniformity of Dosage Units 905:
Meet the requirements
IMPURITIES
Organic Impurities
• Procedure
Mobile phase, Standard solution, and Sample solution:
Prepare as directed in the Assay.
System suitability solution:
0.025 mg/mL of USP Orlistat Related Compound D RS in Mobile phase. Transfer 1 mL of this solution to a 50-mL volumetric flask, and dilute with Standard solution to volume.
Chromatographic system
System suitability
Sample:
System suitability solution
System suitability requirements
Resolution:
NLT 1.4 between USP Orlistat RS and USP Orlistat Related Compound D RS
Relative standard deviation:
NMT 2.0% for the orlistat peak
Analysis
Samples:
Standard solution and Sample solution
Calculate the percentage of each impurity in the portion of Capsules taken:
Result = (rU/rS) × (CS/CU) × (1/F) × 100
Acceptance criteria:
See Impurity Table 1.
Impurity Table 1
ADDITIONAL REQUIREMENTS
• Packaging and Storage:
Preserve in tight containers, and store at 25, excursions permitted between 15 and 30.
USP35
Auxiliary Information
Please check for your question in the FAQs before contacting USP.
USP35NF30 Page 4133
Pharmacopeial Forum: Volume No. 35(5) Page 1169
|