|
Mercaptopurine Tablets
DEFINITION
Mercaptopurine Tablets contain NLT 93.0% and NMT 110.0% of the labeled amount of mercaptopurine (C5H4N4S·H2O).
IDENTIFICATION
Change to read:
• A.
The UV absorption spectrum exhibits a maximum at 325 ± 2 nm, and the ratio A255/A325 does not exceed 0.09.
Sample:
Change to read:
• B.
ASSAY
Change to read:
• Procedure
Solution B:
Methanol and Solution A (5:95)
Solution C:
Methanol and Solution A (30:70)
Mobile phase:
Solution B and Solution C (80:20)
Diluent:
Methanol and water (1:1)
Standard solution:
0.25 mg/mL of USP Mercaptopurine RS in a mixture of methanol and water (1:1). Transfer USP Mercaptopurine RS into a suitable volumetric flask, and add methanol equivalent to 50% of the final volume. Shake mechanically to dissolve, and dilute with water to volume.
Sample stock solution:
0.5 mg/mL of mercaptopurine in Diluent from NLT 5 Tablets. Place the Tablets into a suitable volumetric flask, add methanol equivalent to 50% of the final volume, and shake mechanically for a minimum of 30 min. Dilute with water to volume. Pass through a PVDF filter of 0.45-µm pore size, and discard the first 3 mL of filtrate.
Sample solution:
0.25 mg/mL of mercaptopurine in Diluent from the Sample stock solution
Chromatographic system
Mode:
LC
Detector:
UV 260 nm
Column:
4.6-mm × 10-cm; 3-µm packing L1
Column temperature:
30
Flow rate:
1.0 mL/min
Injection size:
10 µL
System suitability
Sample:
Standard solution
Suitability requirements
Tailing factor:
NMT 2.0
Relative standard deviation:
NMT 2.0%
Analysis
Samples:
Standard solution and Sample solution
Calculate the percentage of mercaptopurine (C5H4N4S·H2O) in the portion of Tablets taken:
Result = (rU/rS) × (CS/CU) × (Mr1/Mr2) × 100
Acceptance criteria:
93.0%110.0%
PERFORMANCE TESTS
Change to read:
• Dissolution
Test 1
Medium:
0.1 N hydrochloric acid; 900 mL
Apparatus 2:
50 rpm
Time:
60 min
Mobile phase:
0.1% acetic acid in water
Standard solution:
USP Mercaptopurine RS in Medium
Sample solution:
Pass a portion of the solution under test through a suitable filter. Dilute with Medium to a concentration that is similar to the Standard solution, if necessary.
Chromatographic system
Mode:
LC
Detector:
UV 230 nm
Column:
3.9-mm × 15-cm; packing L1
Flow rate:
2.5 mL/min
Injection size:
10 µL
System suitability
Sample:
Standard solution
[NoteThe retention time for mercaptopurine is NLT 4 min. ]
Suitability requirements
Relative standard deviation:
NMT 2.0%
Analysis
Samples:
Standard solution and Sample solution
Tolerances:
NLT 80% (Q) of the labeled amount of C5H4N4S·
Test 2:
If the product complies with this test, the labeling indicates that it meets USP Dissolution Test 2.
Medium, Apparatus 2, Chromatographic system, and Analysis:
Proceed as directed for Test 1.
Time:
120 min
Tolerances
NLT 80% (Q) of the labeled amount of C5H4N4S·H2O is dissolved.
• Uniformity of Dosage Units
IMPURITIES
Add the following:
• Procedure
Solution A:
0.1% (v/v) formic acid in water
Solution B:
Methanol and Solution A (2:98)
Solution C:
Methanol and Solution A (1:1)
Mobile phase:
See Table 1.
Table 1
Standard stock solution:
0.06 mg/mL of USP Mercaptopurine RS in Solution A. [NoteUse methanol equivalent to 2.5% of the final volume to help dissolve. ]
Standard solution:
1.2 µg/mL of USP Mercaptopurine RS in Solution B from the Standard stock solution
Sensitivity solution:
0.06 µg/mL of USP Mercaptopurine RS in Solution B from the Standard solution
Sample stock solution:
0.5 mg/mL of mercaptopurine in a mixture of methanol and Solution A (1:9) from NLT 5 Tablets. Place the Tablets into a suitable volumetric flask, add methanol equivalent to 10% of the final volume, and shake mechanically for a minimum of 30 min. Dilute with water to volume. Pass through a PVDF filter of 0.45-µm pore size, and discard the first 3 mL of filtrate.
Sample solution:
0.12 mg/mL of mercaptopurine in Solution A. Transfer 6.0 mL of the Sample stock solution into a 25-mL volumetric flask, and dilute with Solution A to volume. Pass through a PVDF filter of 0.45-µm pore size, and discard the first 5 mL of filtrate. [NoteInject the Sample solution within 1 h of preparation. ]
Chromatographic system
Mode:
LC
Detector:
UV 260 nm
Column:
4.6-mm × 10-cm; 3-µm packing L1
Temperature
Column:
30
Sample:
4
Flow rate:
1.0 mL/min
Injection size:
50 µL
System suitability
Samples:
Standard solution and Sensitivity solution
Suitability requirements
Tailing factor:
NMT 2.0, Standard solution
Signal-to-noise ratio:
NLT 10, Sensitivity solution
Relative standard deviation:
NMT 2.0%, Standard solution
Analysis
Samples:
Standard solution and Sample solution
Calculate the percentage of each impurity in the portion of Tablets taken:
Result = (rU/rS) × (CS/CU) × (1/F) × 100
Acceptance criteria
Individual impurities:
See Table 2. [NoteDisregard any impurity peak less than 0.05%. ]
Table 2
ADDITIONAL REQUIREMENTS
• Packaging and Storage:
Preserve in well-closed containers.
• Labeling:
When more than one Dissolution test is given, the labeling states the Dissolution test used only if Test 1 is not used.
Auxiliary Information
Please check for your question in the FAQs before contacting USP.
USP35NF30 Page 3811
Pharmacopeial Forum: Volume No. 36(6) Page 1554
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||