Minocycline Periodontal System
DEFINITION
Minocycline Periodontal System is an extended-release formulation of Minocycline Hydrochloride containing the equivalent of NLT 90.0% and NMT 120.0% of the labeled amount of minocycline (C23H27N3O7).
IDENTIFICATION
•  A. Ultraviolet Absorption 197U
Wavelength range:  250–450 nm
Standard stock solution:  Transfer USP Minocycline Hydrochloride RS to a suitable volumetric flask. Dissolve first in dimethylformamide, using about 20% of the final volume, then dilute with water to volume, and mix to obtain a solution having a known concentration of about 0.48 mg/mL of minocycline.
Standard solution:  0.024 mg/mL minocycline hydrochloride from Standard stock solution in water
Sample stock solution:  Transfer Minocycline Periodontal System equivalent to 12 mg of minocycline hydrochloride to a 25-mL volumetric flask. Add 5.0 mL of dimethylformamide, and mix to dissolve. Dilute with water to volume and filter.
Sample solution:  0.024 mg/mL minocycline hydrochloride from Sample stock solution in water
Acceptance criteria:  The Sample solution exhibits maxima at the same wavelengths as the Standard solution, concomitantly measured.
•  B. The retention time of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.
ASSAY
•  Procedure
Mobile phase:  Dimethylformamide, 0.2 M ammonium oxalate, and 0.1 M edetate disodium (25:55:20), adjusted with 0.4 M aqueous tetrabutylammonium hydroxide solution to a pH of 6.3 ± 0.2
Diluent:  Dimethylformamide and methanol (1:1)
System suitability solution:  Prepare a solution in water containing 2 mg/mL USP Minocycline Hydrochloride RS. Heat over a steam bath for 60 min. To one part of this solution add four parts of Mobile phase and mix. Refrigerate the solution immediately after preparation and during analysis, using a refrigerated autosampler.
Standard solution:  0.4 mg/mL of minocycline from USP Minocycline Hydrochloride RS in Diluent. Refrigerate the solution immediately after preparation and during analysis, using a refrigerated autosampler. [Note—Use low-actinic glassware. ]
Sample solution:  Mix the contents of NLT 10 dispensing units of Minocycline Periodontal System. Transfer a portion of the mixture, equivalent to 10 mg of minocycline, into a 25-mL volumetric flask. Add Diluent and sonicate for 2–5 min, or until the sample is dissolved. Dilute with Diluent to volume, and mix to obtain a solution having a nominal concentration of 0.4 mg/mL of minocycline, based on the label claim. Refrigerate the solution immediately after preparation and during analysis, using a refrigerated autosampler. [Note—Use low-actinic glassware. ]
Chromatographic system 
Mode:  LC
Detector:  UV 280 nm
Guard column:  4.6-mm × 3-cm; 10-µm packing L7
Column:  4.6-mm × 15-cm; 5-µm packing L7
Flow rate:  2 mL/min
Autosampler temperature:  5
Injection size:  10 µL
System suitability 
Samples:  System suitability solution and Standard solution
[Note—The relative retention times of epiminocycline and minocycline are 0.81 and 1, respectively. ]
Suitability requirements 
Resolution:  NLT 2.0 between epiminocycline and minocycline, System suitability solution
Tailing factor:  NMT 2.0 for the minocycline peak, System suitability solution
Relative standard deviation:  NMT 2.0% for the minocycline peak, Standard solution
Analysis 
Samples:  Standard solution and Sample solution
Calculate the percentage of minocycline (C23H27N3O7) in the portion of the Minocycline Periodontal System taken:
Result = (rU/rS) × (CS/CU) × P × F × 100
rU== peak response of minocycline from the Sample solution
rS== peak response of minocycline from the Standard solution
CS== concentration of USP Minocycline Hydrochloride RS in the Standard solution (mg/mL)
CU== nominal concentration of minocycline in the Sample solution (mg/mL)
P== potency of minocycline in USP Minocycline Hydrochloride RS (µg/mg)
F== conversion factor, 0.001 mg/µg
Acceptance criteria:  90.0%–120.0%
PERFORMANCE TESTS
•  Dissolution
Medium:  6.9 g/L of monobasic sodium phosphate monohydrate in water, adjusted with phosphoric acid to a pH of 4.2. This solution is stable for 10 days.
Apparatus:  Tube rotator. [Note—Suitable equipment is available as Labquake® tube rotator, catalog number 400110. ]
0.1 M Edetate disodium:  37.2 g/L of edetate disodium in water
0.2 M Ammonium oxalate:  28.4 g/L of ammonium oxalate in water
Mobile phase:  Mix 310 mL of 0.1 M Edetate disodium and 500 mL of 0.2 M Ammonium oxalate, adjust with 0.4 M aqueous tetrabutylammonium hydroxide to a pH of 6.2, and add 175 mL of dimethylformamide. The injector wash solution is a mixture of dimethylformamide and water (25:75).
Standard stock solution:  0.11 mg/mL of USP Minocycline Hydrochloride RS in Medium
Standard solutions:  Dilute the Standard stock solution with Medium to obtain solutions with final concentrations of 0.088 mg/mL, 0.0528 mg/mL, 0.0352 mg/mL, 0.022 mg/mL, and 0.0176 mg/mL.
System suitability solution:  Transfer 10 mg of USP Minocycline Hydrochloride RS to a 50-mL beaker. Add 5 mL of water and heat on a steam bath for 60 min. Add 20 mL of Medium or Mobile phase, and mix well. Store at 5.
Sample solution:  Use borosilicate glass tubes, 25 mm outside diameter and 15 cm long. Close the tubes with a snap type cell with a Teflon prong consisting of a Teflon closure and holder that snap together, two 25-µm stainless steel screens, two silicone gaskets, and a Teflon spacer (see Figure 1). Prepare six tubes as follows: partially assemble a release tube and tare its weight; dispense one dose of Minocycline Periodontal System into a partially assembled release cell (see Figure 1); record the sample weight in mg; assemble the cell so that the sample is enclosed between the two 25-µm screens; close the cells and place each one of them into separate glass tubes containing 10 mL of Medium previously equilibrated at 37; add the Teflon prong, and cap the tube with Teflon faced rubber-lined caps; seal with Teflon tape. Place the tubes in the tube rotator. Place the tube rotator in a convection incubator that is maintained at 37. Allow the tubes to rotate for 4 h. Remove the solution under test, and add 10 mL of Medium previously equilibrated at 37. Replace the tubes in the apparatus and rotate for 20 h (24 h total). Repeat the sampling procedure after 24 h (48 h total), and after another 24 h (72 h total).
Click to View Image
Figure 1. Sample Extraction Configuration
Chromatographic system 
Mode:  LC
Detector:  UV 280 nm
Guard column:  4.6-mm × 3-cm; 10-µm packing L7
Column:  4.6-mm × 3.3-cm; 5-µm packing L1
Flow rate:  1.5 mL/min
Autosampler temperature:  5
Injection size:  20 µL for the 4 and 24 h time points; 50 µL for the 48 and 72 h time points
Suitability requirements 
Samples:  System suitability solution and Standard solutions
Resolution:  NLT 2.0 between epiminocycline and minocycline. Inject 20 µL of the System suitability solution.
Tailing factor:  NMT 2.0. Inject 20 µL of the System suitability solution.
Relative standard deviation:  NMT 2.0% for the minocycline peak, any of the Standard solutions
Analysis:  Construct a calibration curve for each sampling interval by plotting the concentration of the Standard solutions versus peak area. Calculate the slopes and y-intercepts using linear regression analysis.
Calculate the release rate of minocycline:
Resulti = [(rUi yi)/Si] × 10/( i × W × A)
i== sampling time, 4, 24, 48, 72 h
rUi== peak response from each of the Standard solutions at time i
yi== y-intercept of the calibration curve at sampling time i
Si == slope of the calibration curve at sampling time i
W== weight of the sample (mg)
A== amount of minocycline in the sample (mg/mg of sample) as determined in the Assay
Tolerances 
Time
(h)
Release Rate (µg/h)
Average of 6 Measurements
0–4 NLT 25
4–24 NLT 1.0
24–48 NLT 0.2
48–72 NLT 0.05
•  Uniformity of Dosage Units 905: Meet the requirements
IMPURITIES
Organic Impurities 
•  Procedure
Mobile phase, Diluent, System suitability solution, Standard solution, Sample solution, Chromatographic system, and System suitability:  Proceed as directed in the Assay.
Analysis 
Sample:  Sample solution
Calculate the percentage of each related compound in the portion of Minocycline Periodontal System taken:
Result = (rU/rT) × 100
rU== peak response of each impurity from the Sample solution
rT== sum of the peak responses from the Sample solution. [Note—Exclude peaks eluting in the solvent front. ]
Acceptance criteria 
Individual impurities:  NMT 6.0% of epiminocycline
Total impurities (excluding epiminocycline):  NMT 3.5%
SPECIFIC TESTS
•  Microbial Enumeration Tests 61 and Tests for Specified Organisms 62: The total aerobic microbial count does not exceed 1000 cfu/g; the total combined molds and yeasts count does not exceed 100 cfu/g; and the product meets the requirements of the test for the absence of Escherichia coli.
ADDITIONAL REQUIREMENTS
•  Packaging and Storage: Preserve in a tight, light-resistant container. Store at controlled room temperature.
•  USP Reference Standards 11
USP Minocycline Hydrochloride RS Click to View Structure
Auxiliary Information— Please check for your question in the FAQs before contacting USP.
Topic/Question Contact Expert Committee
Monograph Ahalya Wise, M.S.
Senior Scientific Liaison
1-301-816-8161
(SM12010) Monographs - Small Molecules 1
61 Radhakrishna S Tirumalai, Ph.D.
Principal Scientific Liaison
1-301-816-8339
(GCM2010) General Chapters - Microbiology
62 Radhakrishna S Tirumalai, Ph.D.
Principal Scientific Liaison
1-301-816-8339
(GCM2010) General Chapters - Microbiology
Reference Standards RS Technical Services
1-301-816-8129
rstech@usp.org
USP35–NF30 Page 3929
Pharmacopeial Forum: Volume No. 36(2) Page 420