|
Add the following:
This chapter provides procedures used to determine molecular weight (MW) distribution and weight-average molecular weight for low molecular weight heparins (LMWH).
INTRODUCTION
Low molecular weight heparins are prepared from Heparin Sodium, USP, by partial depolymerization. LMWHs are polydisperse, i.e., they are made up of polysaccharide chains with a range of molecular weights. The MW distribution is a defining characteristic for each LMWH product. The USP contains specific monographs for LMWH products, including limits on molecular weight distribution parameters.
Most techniques for determination of the MW distribution of LMWHs make use of gel permeation chromatography (GPC), sometimes referred to as size exclusion chromatography (SEC). In order to derive molecular weight distribution from a chromatogram, it is necessary to know the relationship between retention time and MW. This chapter describes a GPC method that is calibrated using the USP Low Molecular Weight Heparin Molecular Weight Calibrant Reference Standard (RS). This preparation is a single polydisperse Standard for which the distribution of molecular weights is described in the form of a Broad Standard Table (see the USP Certificate for USP Low Molecular Weight Heparin Molecular Weight Calibrant RS) that lists the log of the molecular weight (log MW) of a number of reference points and the matching percent fractions by weight of the calibrant above or below these reference points. The Broad Standard Table is used together with a chromatogram of the calibrant to fit a suitable function, in this case a third degree polynomial, to the relationship between log MW and retention time, thus generating a calibration curve for the chromatographic system. With this calibration curve, the analyst calculates the weight-average molecular weight and distribution parameters from the chromatogram of a sample of LMWH.
All the calculations described in this chapter may be performed using a spreadsheet program, but this method is not recommended as it is laborious and open to human error. Proprietary software capable of automating the calibration and molecular weight calculations can be obtained from a number of chromatography systems manufacturers.
PROCEDURE
• Molecular Weight Measurements of Low Molecular Weight Heparins by Gel Permeation Chromatography
The following procedure, with any necessary variations, is used where specified in the individual monographs.
Ammonium acetate stock solution:
Sodium azide solution:
Mobile phase:
Calibration solution:
Sample solution:
System suitability solution:
Chromatographic system
[NoteThe temperature of the refractive index detector must be set at the same temperature as that of the Column temperature.
Mode:
Detector:
Columns
Analytical:
Guard:
Column temperature:
Flow rate:
Column equilibration:
Injection volume:
System suitability
Sample:
Suitability requirements
Resolution:
Calibration curve:
Weight-average molecular weight (Mw):
Analysis
Samples:
[NoteThe calibrant, Standard, or sample of low molecular weight heparin will give a broad peak between about 25 and 45 min, followed by a later eluting narrow salt peak, as illustrated in the USP Certificate for USP Low Molecular Weight Heparin Molecular Weight Calibrant RS.
Calculations:
Using the same GPC software, for each of the duplicate chromatograms of the System suitability solution and the Sample solution, with the calibration function derived as described above, calculate Mw:
Mw = S(RIi × Mi)/SRIi
RIi = detector response at each point Mi = MW at each point Round the mean value of Mw to the nearest 50 Da. Using the same GPC software, determine for each of the duplicate Sample solution chromatograms the percentage of product with molecular weight as indicated in the product monograph.
Acceptance criteria:
ADDITIONAL REQUIREMENTS
• USP Reference Standards
USP Dalteparin Sodium RS
USP Low Molecular Weight Heparin Molecular Weight Calibrant RS
1
The method was validated using a guard column TSK SWXL 6-mm × 4-mm; 7-µm in series with two analytical columns: TSK G3000 SWXL 7.8-mm × 30-cm; 5-µm in series with a TSK G2000 SWXL 7.8-mm × 30-cm; 5-µm.
Auxiliary Information
Please check for your question in the FAQs before contacting USP.
USP38NF33 Page 232
Pharmacopeial Forum: Volume No. 40(1)
|