Acetaminophen Extended-Release Tablets
» Acetaminophen Extended-Release Tablets contain not less than 90.0 percent and not more than 110.0 percent of the labeled amount of acetaminophen (C8H9NO2).
Packaging and storage— Preserve in tight containers.
Labeling— Where the Tablets are gelatin-coated, the label so states. When more than one Dissolution test is given, the labeling states the Dissolution test used only if Test 1 is not used.
Identification—
A: Infrared Absorption 197K—Use a portion of powdered Tablets.
B: The retention time of the major peak in the chromatogram of the Assay preparation corresponds to that in the chromatogram of the Standard preparation, as obtained in the Assay.
Dissolution 711
test 1
Medium: simulated gastric fluid TS (without enzyme); 900 mL.
Apparatus 2: 50 rpm.
Times: 15 minutes, 1 hour, and 3 hours.
Procedure— Determine the amount of C8H9NO2 dissolved from UV absorbances at 280 nm, using a filtered portion of the solution under test in comparison with a Standard solution having a known concentration of USP Acetaminophen RS in the same Medium.
Tolerances— The percentages of the labeled amount of C8H9NO2 dissolved at the times specified conform to Acceptance Table 2.
Time Amount dissolved
15 minutes between 45% and 65%
1 hour between 60% and 85%
3 hours not less than 85%
for gelatin-coated tablets—
Medium, Apparatus, and Procedure— Proceed as directed above.
Times: 30 minutes, 90 minutes, and 4 hours.
Tolerances— The percentage of the labeled amount of C8H9NO2 dissolved at the times specified conform to Acceptance Table 2.
Time Amount dissolved
30 minutes between 40% and 60%
90 minutes between 55% and 85%
4 hours not less than 80%
test 2— If the product complies with this test, the labeling indicates that it meets USP Dissolution Test 2.
Medium, Apparatus, and Procedure— Proceed as directed for Test 1.
Tolerances— The percentages of the labeled amount of C8H9NO2 dissolved at the times specified conform to Acceptance Table 2.
Time Amount dissolved
15 minutes between 40% and 60%
1 hour between 55% and 75%
3 hours not less than 80%
Uniformity of dosage units 905: meet the requirements.
Assay—
Mobile phase— Prepare a mixture of water and phosphoric acid (9:1). Combine 1 mL of this solution with a mixture of water and methanol (700:300). Filter and degas. Make adjustments if necessary (see System Suitability under Chromatography 621).
Standard preparation— Dissolve an accurately weighed quantity of USP Acetaminophen RS in methanol, and dilute quantitatively, and stepwise if necessary, with Mobile phase to obtain a solution having a known concentration of about 0.65 mg per mL.
Assay preparation— Transfer 10 Tablets into a 250-mL volumetric flask containing 50 mL of water and a magnetic stir bar. Stir at least 30 minutes or until the coating has dissolved. Add 150 mL of methanol, and stir for 45 minutes. Tablet cores should be disintegrated at least 15 minutes prior to ending the stirring. Remove the magnetic stir bar and rinse into the flask with methanol. Dilute with methanol to volume, mix well, and centrifuge. Transfer 5 mL of the clear supernatant to a 200-mL volumetric flask, dilute with Mobile phase to volume, and mix well.
Chromatographic system (see Chromatography 621)— The liquid chromatograph is equipped with a 295-nm detector and a 3.9-mm × 15-cm column that contains packing L1. The flow rate is about 2.0 mL per minute. Chromatograph the Standard preparation, and record the peak responses as directed for Procedure: the tailing factor is not more than 3.0, and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure— Separately inject equal volumes (about 20 µL) of the Standard preparation and the Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of acetaminophen (C8H9NO2) in each Tablet taken by the formula:
1000C(rU / rS)
in which C is the concentration, in mg per mL, of USP Acetaminophen RS in the Standard preparation; and rU and rS are the acetaminophen peak responses obtained from the Assay preparation and the Standard preparation, respectively.
Auxiliary Information— Please check for your question in the FAQs before contacting USP.
Topic/Question Contact Expert Committee
Monograph Clydewyn M. Anthony, Ph.D.
Scientist
1-301-816-8139
(MDCCA05) Monograph Development-Cough Cold and Analgesics
Reference Standards Lili Wang, Technical Services Scientist
1-301-816-8129
RSTech@usp.org
711 Margareth R.C. Marques, Ph.D.
Senior Scientist
1-301-816-8106
(BPC05) Biopharmaceutics05
USP32–NF27 Page 1391
Pharmacopeial Forum: Volume No. 32(6) Page 1666
Chromatographic Column—
Chromatographic columns text is not derived from, and not part of, USP 32 or NF 27.